Leica GP2 (SOP)
Leica GP2 (SOP)
Before using the plunger, any new user has to undergo a standard training. The goal of this training is for the user to become independent enough to do the grid plunging by themselves. The facility staff will decide at which moment this happens. Once independent, the user can work in out of business hours without observation. However, every user is free to ask at any time for help and advice from the facility staff.
The plunger needs to be booked through the booking system. And if no one wants to use the machine directly after you, you have to run a bake out cycle which will last at least 70 minutes.
Download file GP2_Operating_manual.pdf
View file GP2_Operating_manual.pdf
GP2 operating manual by Leica |
plunger preparation
- Touch the screen to initialize the plunger (Figure 1).
- After initialization a warning appears to inform you about the preparation steps upcoming (Figure 2).
- Reinsert the drip tray on the right side of the plunger (Figure 3).
- Replace the filter in the plunging chamber and attach it with the magnetic disc (Figure 4&5). Important: The machine is only calibrated for Whatman 1 paper.
- Attach the water reservoir on the right side of the plunger (Figure 6) and tighten it with the screw (Figure 7).
- Add 70 ml MilliQ water through the 3 way valve (Figure 8).
- Add the black ethane container in the cooling area (Figure 9).
- Add the grid box storage container into the cooling area (Figure 9).
- Cool and fill the area with LN2 until just above the metal mesh (Figure 9). You will need around 1.8 l.
- Click 'OK' on the touchscreen (Figure 2)
- To condense the ethane attach the secondary cryogen liquefier onto the ethane container (Figure 10).
- Connect the hose with the connector to the ethane gas tank (Figure 11).
- Open the main valve of the ethane tank.
- Open the regulator valve of the ethane tank until you see gas entering the liquefier.
- Ethane will condense in the chamber and you can see the ethane level due to surface reflections (Figure 12).
- Once the ethane container is filled, close the regulator valve and the main valve of the ethane bottle.
- Slowly lift up the liquefier and make sure to not drop excess ethane around.
- Important: Make sure that the outer part of the cryo area can still move down and is not frozen in place
- Top up the liquid nitrogen until the mesh is just covered.
- Note: Make sure the ethane is cooled before starting the plunging. At this point you may want to do the glow discharging of your grid.
Front side blotting
We have two stored programs for front side blotting. One at 6 degrees Celsius and one at 20 degrees. Here the sample blotted on the same side as it was applied.
The GP2 is only calibrated for Whatman 1 paper and per design expects 4 ul of sample due to the setup of the blotting sensor. 3 ul sample seems to work always as well. 2 ul sample needs manual intervention in 50% of the cases, but can be done.
- In the main display you can see the chamber temperature, relative humidity, filling level of LN2 and available blotting positions (Figure 1)
- If you have changed the blotting paper click on the two circling arrows next to 'blot' to reset the counter to 10 (Figure 1).
- On the main display click on 'Programs' and select the program (either 1 or 2) called "frontside ..." (Figure 2)
- Note: Give the chamber enough time to cool to 6 degrees (or heat) to reach equilibrium, before starting with your first grid.
- On the home screen (Figure 3) you have the chance to change some parameters of your grid plunging procedure.
- By clicking on the pencil icon next to chamber on the home screen (Figure 3, top left), you can open the display for the environment (Figure 4) to adapt the chamber temperature and the humidity.
- For this click on 'Set' for either option (Figure 4) and type in the desired temperature (Figure 5) or humidity (Figure 6).
- Confirm by clicking 'OK' (Figure 5/6)
- Either go back to home screen by clicking 'Main' or to the 'load specimen' screen by clicking that button (Figure 4).
- Note: changing the humidity is not recommended
- Note 2: changing the temperature will required equilibration time.
- By clicking on the pencil icon next to load specimen on the home screen (Figure 3), you can open the display for specimen loading (Figure 7) to adapt incubation time before blotting.
- For this click on 'Set' next to delay time and adjust the time (Figure 7&8)
- Confirm with 'OK' (Figure 8)
- Either go back to home screen by clicking 'Main' or to the 'Blot' screen by clicking that button (Figure 4).
- By clicking on the pencil icon next to 'Blot' on the home screen (Figure 3), you can open the display for Blotting (Figure 9) to adapt blotting time.
- For this click on 'Set' next to Blot time and adjust the time (Figure 10)
- Confirm with 'OK' (Figure 10)
- Either go back to home screen by clicking 'Main' or to the 'Plunge/Transfer' screen by clicking that button (Figure 4).
- By clicking on the pencil icon next to 'Plunge' or 'Transfer' on the home screen (Figure 3), you can open the display for Plunge/Transfer (Figure 11) to adapt the post blot delay.
- For this click on 'Set' next to Post-blotting time and adjust the time (Figure 11&12)
- Confirm with 'OK' (Figure 12)
- Go back to home screen by clicking 'Main' (Figure 4).
- On the main display click on 'Load forceps' (or use the foot pedal). The plunging rod will move (Figure 13).
- Pick up a grid with the plunging tweezer and lock it with the black sleeve. Ensure proper orientation.
- Important: The picture below shows the top side of the grid. When the 'H' on the tweezer is facing you, the top side for sample application needs to face to the left!
- Load the tweezer onto the gantry of the plunger (Figure 15).
- Note: The 'H' symbol of the tweezer will point towards you
- Push the foot pedal to lower the chamber and lift the tweezer.
- Note: The tweezer will automatically rotate, so that the top side will be facing to the right.
- Apply you sample through the access hole on the right (Figure 16).
- Push the foot pedal.
- The tweezer will turn automatically and the plunging process will commence.
- Error Note: If you use less than 3 ul of sample, the blotting sensor may not detect the blotting. This results in a warning (Figure 17). Click 'OK' and push immediately the 'Plunge' button. Grid can still be used!
- After plunging the tweezer can be removed from the gantry, while staying in the cold area
- Hold the grid with the tweezer and loosen the black sleeve.
- Optional: Touch your grid on a filter paper that you placed in the LN2 chamber to wick away excess ethane (Figure 18).
- Transfer your grid into the grid box.
- Push the foot pedal to lift the gantry (Figure 19).
- Repeat for the other grids while varying the plunging time.
- Note: The next foot pedal push will move the gantry into the tweezer loading position.
- After four grids are plunged and added to your grid box, close the grid box
- Make sure the is still LN2 in the Cryo transfer container.
- Put the lid onto the cryo transfer container (Figure 20).
- Screw the M4 cryotool to the transfer container (Figure 21).
- Move your grid box with the cryo transfer container into a styrofoam Dewar for transfer into the grid storage system.
- Once finished with your sample, continue with the bake out procedure.
Back side blotting
- Select the program
Hanging drop blotting
- Select the program
Multiple Application
- Select the program
Bake out cycle
The bake out should be run after every use and takes around 70 minutes to complete
- In the main window click on 'Settings' (bottom left, Figure 1).
- In the settings window, click on 'bake out' (Figure 2).
- A warning message appears (Figure 3).
- Remove the ethane vessel and the cryo transfer container (if not done yet during grid transfer) The plunging area should look like Figure 4.
- A new warning appears (Figure 5)
- Remove the filter paper and the magnetic disc. The chamber should look as in Figure 6.
- Another warning appears (Figure 7). Note: We do not have a viewing system.
- Open the front door of the blotting chamber as in Figure 8.
- On the Panel click Start to begin the bake out (Figure 9).
- The timer will only start once the liquid nitrogen is evaporated and the temperature reached 100 degrees (Figure 10).
- Remove the humidifier reservoir by loosening the screw (Figure 11) and carefully moving down with the container (Figure 12)
- Remove the excess liquid.
- Dry the chamber with paper.
- Remove the dripping tray (Figure 13).
- Remove access liquid.
- Dry the dripping tray with paper.
- Wipe off access liquid from the nozzles at the humidifier reservoir with paper (Figure 13)
- Remove big puddles of water in the blotting chamber with paper.
- Put all Leica parts back into the box as shown in Figure 14.