Post Bake Out
POST BAKE OUT
- HT Conditioning/restart/cooling
- GunAlignment
Cooling and conditioning after bake out
- Start refilling of both Liquid nitrogen tanks by clicking “Auto” in the “LN2 Auto Refill Maintenance window” for “transfer tank” and “stage tank”.
- In TEM center on the microscope pictogram click on “HT”.
- In the operation for “Conditioning Rod” click on “Condition”.
- Once the rod is in position (pink progress bar), click in the line “HT” the button “ON”
- Wait until procedure is finished (~55 minutes).
- If you realize that one of the nitrogen tanks gets refilled start manually also the refill of the second tank by clicking “Refill” in the “LN2 Auto Refill Maintenance” window.
- Once finished HT will be back to 0 kV.
- Switch the “conditioning Rod” back to “Operate”.
- Click on “HT” the button “ON”
- Wait until procedure is finished (~25 minutes).
- close HT window
- In the microscope pictogram hover over “Emission” and click on “Flashing” the button “High”.
- When done click in the same window on “Auto Emission” “ON”.
- Emission will ramp up to 10 uA.
Gun Alignment
DO NOT DO THIS EVEN AS A USER OR SUPERUSER!!!
- Open the beam valve
- Switch off Low Dose mode in SerialEM (if it is on).
- Go to Spot size 1.
- Set Beam Align 1 to Default.
- Go to 10k or 20k Magnification
- Set exposure time on large phosphor screen to 0.001s (minimum)
- Remove CL aperture.
- Condense beam.
- Center beam with GunAlign 1. Focus on the bright spot as center.
- Switch on Anode wobbler.
- Make beam concentric between the changes using GunAlign2. (Spreading in all direction equivalently).
- Repeat GunAlign1 and Anode wobbler until you do not have to reposition GunAlign1 anymore.
- Go to the Spot size you will be working with (i.e. 6)
- Go to 60k Magnification.
- Insert CL aperture.
- Set exposure time on large phosphor screen to 0.02 s.
- Switch on Anode wobbler.
- Adjust GunAlign2 so that the beam is concentric between the changes.
- Control the concentricity with CL aperture removed (Above procedure only works well if CL aperture is centered). Adjust GunAlign2 if necessary.